Mako R/F Probe

cart 0
0
Product Page
Mako R/F Probe
RTI

Mako R/F Probe

Enquiry Item

The Mako R/F Probe, paired with the Mako Base Unit, offers industry-leading accuracy (±1.5% kV measurement uncertainty) and exceptional sensitivity, covering the full range of dose rates.

Its unique design ensures easy setup in the X-ray beam, flexible connections between the Probe and Base Unit, and a small radiological footprint—perfect for seamless measurements without disrupting the AEC system.

Ideal for Radiography, Fluoroscopy, CT, and Dental (Intraoral & Panoramic) applications.

Qty


Automated, Data-Driven Excellence

Step into a world where data collection is streamlined and efficient. Ocean Next™ offers automated routines and complete traceability, with real-time data gathering from Mako and its probes. Pre-defined templates ensure smooth collaboration and quick setup, keeping you ahead of the curve.

The Future-Proof Solution

Mako is designed to evolve with your needs. Its multi-modal design and continuous innovation mean it will always adapt to include new functions, probes, and expanded applications, ensuring it remains your go-to tool for the future.

All-in-One Measurement Solution

Mako is more than just a tool—it's a complete solution. Compact, cable-free, and designed for multiple uses, it handles all the key parameters you need. With its high sensitivity, Mako delivers quick and accurate measurements, providing precise results faster than other tools.

General Speicifications

  • Size: 122.5 x 28 x 14.5 mm (4.82 x 1.1 x 0.57 in)
  • Weight: 55 g (1.94 oz)
  • Detector active areas:
    • Smallest 0.9 x 0.9 mm2
    • Largest 7 x 7 mm2
  • Detector position check: Orientation independent, no check required
  • Backscatter Protection: Yes


Measuring Specifications (Mako R/F)

  • Tube Potential
    • Range: 35 – 155 kV
    • Min dose rate: 2 µGy/s
    • PPV and kVp: Yes
    • Uncertainty: ±1.5 %
  • Dose
    • Range: 1 nGy – 9999 Gy
    • Uncertainty: ±5 % or 5 nGy
  • Dose Rate
    • Range: 1 nGy/s – 500 mGy/s
    • Resolution: 0.1 nGy/s
    • Trigg level:
      • 25 nGy/s @ exp. time >200 ms
      • 3 µGy/s @ 1 ms
      • 0.3 µGy/s @ 10 ms
    • Uncertainty: ±5 % or 1 nGy/s
  • Time
    • Range 0.33 ms – 48 h
    • Resolution 0.33 ms Bandwidth Up to 1 kHz
    • Uncertainty: ±0.5 % or 0.33 ms
  • HVL
    • Range: 1 – 15 mm Al
    • Min dose rate: 0.5 µGy/s
    • Uncertainty: ±5 % or 0.2 mm Al
  • Total Filtration
    • Range 1 – 90 mm Al equivalent
    • Min dose rate 2 µGy/s
    • Uncertainty ±10 % or 0.2 mm Al
  • Pulses
    • Range 1 – 5 000 000
    • Min peak dose rate 0.25 µGy/s
    • Min pulse width 0.33 ms
  • Pulse Rate
    • Range 0.2 – 260 Hz
    • Min peak dose rate 0.25 µGy/s
    • Min pulse width 0.33 ms
  • Waveforms
    • Range 0 – 240 s Resolution 0.33 ms at exposures <15 s
    • Bandwidth kV 1 kHz
    • Bandwidth Dose Rate 1 kHz