ECG100S

cart 0
0
Product Page
ECG100S
Cardioline

Cardioline ECG100S

Enquiry Item

The 12 lead ECG for your department

Plus full connectivity (LAN/WiFi). Telemedicine ready

An effective and powerful 12 lead Diagnostic Electrocardiograph, optimizing ECG workflow with the help of a smart user interface and connectivity to dedicated ECG management systems.

The ECG 100S is equipped with quick ECG acquisition for a fast and easy operation, providing highly accurate measurements exceeding the most stringent standards and international recommendations.

Qty


The 12 lead Electrocardiograph for your department. An effective and powerful 12 lead diagnostic electrocardiograph optimizes hospital ECG workflow thanks to an intuitive user interface and seamless data integration.

An effective and powerful 12 lead diagnostic electrocardiograph optimizes ECG workflow thanks to a smart user interface and connectivity to dedicated ECG management systems.

Quick ECG acquisition for fast and easy operation. Highly accurate measurements exceed the most stringent standards and international recommendations. ECG formats: SCP, PDF.

Dedicated networking (LAN, Wi-Fi) to the Cardioline ECGWebApp application, our ECG management system for editing and storage, is totally based on Web technology.

Alternatively, for less demanding applications, the ECG EasyApp has been designed to allow local management of patients’ ECGs from their PC. Glasgow ECG Interpretation Option for Adult and Paediatric Patients.

Optional GDPR module, allows complete protection of sensitive data stored on the electrocardiograph by means of a strong authentication module installed on the device.

  • ECG channels: 12-lead (I, II, III, aVR-L-F, V1-6)
  • Patient Cable: Standard: 15D, 10 wires patient cable
  • CMRR: > 100dB
  • Input impedance: 100MΩ
  • A to D converter: 24 bit, 32000 samples/second/channel
  • Sampling rate of the input stage: 32000 samples/second/channel
  • Sampling rate for signal analysis: 1000 samples/second/channel
  • A/D conversion: 20 bit
  • Output Data Resolution: 1 μV/LSB
  • Dynamic Range: +/- 400 mV
  • Bandwith: 0,05-300 Hz
  • Pacemaker detection: Hardware detection coupled with convolution digital filtering
  • Defibrillation Protection: AAMI/IEC standards
  • Front-end performance: ANSI/AAMI IEC 60601-2-25:2011
  • Acquisition Mode: Automatic (12 leads), Manual (3/6 leads), Stat (12 leads)
  • Lead Configuration: Standard, Cabrera

​​​​​​​