HotDog WC77 Controller
The HOTDOG WC77 Temperature Management Controller is the future of patient warming. This multiport controller offers maximum versatility and efficiency in patient temperature management, bringing the HotDog Patient Warming System to the forefront of advanced, patient-centered care.
Advanced Patient Warming System
The HOTDOG WC77 is the most advanced patient warming system ever developed. It is designed to provide maximum versatility and efficiency in patient temperature management, bringing the HotDog Patient Warming System to the forefront of advanced, patient-centered care. The aesthetics, touch screen interface, communication features, data feedback, and overall functionality are all designed to benefit the clinician and patient. It is compatible with older and current warming blankets, mattresses, and cables.
7-inch touch screen with Intuitive User Interface
The WC77 features a 7-inch touchscreen with an intuitive user interface. This reliable patient warming system is at your fingertips, allowing you to control all devices at once or individually by touch. Patient warming has been simplified so you can focus on other aspects of the procedure.
Temperature Monitoring
The WC77 offers temperature monitoring, displaying patient core temperature on the screen or an external monitor. This feature allows you to keep track of your patient’s temperature from the start to the end of the procedure. Core temperature is always readily available on the interface.
Customizable Graph Display
The WC77 features a customizable graph display, allowing you to zoom and scroll to any time of the procedure to view temperatures at every moment, from beginning to end.
Auto Mode
The temperature monitoring feature allows automatic control and adjustment of the warming devices based on the patient’s temperature.
Powers multiple devices at once
The WC77 can power multiple devices at once. It can use up to 4 warming blankets with ports A (located on the front) through D at one time. Additionally, a warming mattress can be added with Port M to warm above and below simultaneously. All devices can be controlled at once or individually to suit your needs.
Specifications
Physical Characteristics
- Dimensions: 22.2 cm W x 28 cm H x 17.8 cm D / 8.75” W x 11” H x 7” D
- Weight: 3.75 kg (8.3 lbs) without clamp or cable
- Mounting: Can be placed on a horizontal flat surface (i.e. tabletop), clamped to an IV pole, or hung using a VESA mount of either FDMI MIS-C (35 x 75 mm) or FDMI MIS-D (75 x 75 mm) specifications.
- Settings: Adjustable volume, brightness, language, shut-off timer and more.
Temperature Characteristics
- Temperature control: Micro-processor
- Operating Temperatures:
- Blanket Ports A, B, C, and D adjustable in 1°C increments 37° to 43° ± 1.0°C 98.6° to 109.4° ± 1.8°
- Mattress Port M adjustable in 1°C increments 35° to 40° ± 1.0°C 95° to 104° ± 1.8°F
Safety System
(All alarm conditions are classified as Medium Priority Technical Alarms)
- Auditory Alarms: Minimum SPL of 65 dB(A) at 3m (from front of controller with a background SPL not to exceed 55dB(A)
- Primary Over-temp Alarm:
- Ports A, B, C, D (Warming Blanket) Medium Priority Alarm sounds when temperature sensor is at set point + 1°C
- Port M (Warming Mattress) Medium Priority Alarm sounds when temperature sensor is at set point +1°C
- Secondary Over-temp Alarm:
- Ports A, B, C, D (Warming Blanket) Independent electronic circuit shuts the heater off if the Warming Blanket temperature sensor reaches max set point + 3°C (46°C). Medium Priority Alarm sounds
- Port M (Warming Mattress) Independent electronic circuit shuts the heater off if the Warming Mattress temperature sensor reaches max set point + 2.5°C (42.5°C). Medium Priority Alarm sounds
- Ports A, B, C, D (Warming Blanket) Independent electronic circuit shuts the heater off if the Warming Blanket temperature sensor reaches max set point + 3°C (46°C). Medium Priority Alarm sounds
- Over-current Monitoring:
- Port A - 10 amps max
- Port B - 10 amps max
- Port C - 10 amps max
- Port D - 10 amps max
- Port M - 7 amps max
- Port T TruCore - 1 amp max
- System - 14.6 amps max
- Port A - 10 amps max
Medium Priority Alarm sounds in over current condition. System utilizes power rationing when multiple ports are drawing current over system levels.
- System Over-current Protection: Dual input fused lines. Medium Priority Alarm Sounds
Electrical Characteristics
- Leakage Current: Meets UL 60601-1 and IEC 60601-1 requirements for Class I, Type BF equipment
- Power Consumption: 850W Maximum
- Power Cord: 4.6 m (15 ft) – May vary by country and region per local requirements and regulations
- Device Ratings:
- Input: 100-240 VAC, 50/60 Hz, 850VA
- Output A, B, C, D: 48 VDC, 480 VA Max eachOutput M: 336 VA Max
- Input: 100-240 VAC, 50/60 Hz, 850VA
- Environmental Conditions for Transport and Storage:
- Temperature: -20°C to 60°C
- Humidity: 20% to 80%
- Keep Dry
- Environmental Conditions for Use:
- Temperature: 15°C to 25°C
- Humidity: 20% to 80%
- Fuses: T10AL250V (2 x 5x20mm)