An innovative and broad-spectrum diagnostic lung imaging technology for visualisation of pulmonary ventilation and lung function.

Qty


Indication and usage

Carbon Crucible, when used with sodium pertechnetate Tc 99m in the TechnegasPlus Technegas Generator, provides technetium Tc 99m labelled carbon inhalation aerosol (Technegas®), a radioactive diagnostic agent for use in adults and pediatric patients aged 6 years and older for:

  • visualisation of pulmonary ventilation
  • evaluation of pulmonary embolism when paired with perfusion imaging

The TechnegasPlus Technegas Generator manufactures and delivers our carbon-labelled Technetium 99m ventilation agent; it is primed for optimal physiological assessments of alveolar spaces, facilitated by SPECT or SPECT/CT imaging. Technegas® is developed to address a growing spectrum of conditions and applications, such as:

  • Pulmonary Embolism (PE)
  • Chronic thromboembolic pulmonary hypertension (CTEPH)
  • Chronic Obstructive Pulmonary Disease (COPD)
  • Asthma
  • Pre-operative quantification
  • Radiotherapy treatment planning
  • Identification of common COVID-19 complications

Features:

  • Remote Technegas delivery switch 
  • In-built simmer plate
  • Automatic high-intensity LED for ease of loading
  • Lead glass safety viewing barrier
  • Integrated purge filter to protect personnel
  • Argon gas hose with flow restrictor and quick-release fitting
  • LCD display
  • Lightweight – ergonomic base with 4 locking casters
  • Integrated Patient Administration Set (PAS) shelf
  1. Power
  • Main current:
    • Steady-state <0.2 ampere rms
    • Maximum (17 seconds) 20 ampere rms
  • Installation Requirements:
    • Mains power line*: 200 – 240 +/- 5% VAC
    • Supply frequency: 50 – 60 Hz*
  • Plug Type:
    • North America: NEMA 6-20 P Non-Locking
    • Europe/UK: 32A IP44 2P+E
    • Rest of World: 20A-240V Type I Plug
  1. Environmental
  • Ambient Temperature: 10 – 40 degrees C / 50-104 degrees F
  • Ambient Air Pressure: 70 – 106 kPa / 700 – 1060 mBar
  • Ambient Humidity: 15-90% non-condensing
  1. Unboxed machine dimensions
  • Length: 89 cm / 35 in
  • Width: 50 cm / 19.6 in
  • Height: 137 cm / 53.9 in
  • Weight: 130 kg / 286.6 pounds
  1. Argon & high-pressure regulator
  • Argon purity: High purity greater than 99.996% argon
  • Cylinder size: G or greater
  • Regulator output:
    • Maximum outlet: 150 kpa / 1500 mBar
    • Maximum Flow output: 45 litres/min / 11.88 US gallons/min
  • Regulator connections:
    • Gas bottle connection – as per your local requirements
    • Regulator output: 5/8” UNF RH